S.G, Bé°, C% Tester for Acid Solution TWD-150AS

Flowing liquid testing domain

  • TWD-150AS
    TWD-150AS

 

Application:
Hydrochloric acid, nitric acid, sulfuric acid, phosphoric acid, chromic acid, acetic acid, formic acid, oxalic acid, citric acid, tartaric acid and other production plants, research and experimental units.
 
Principle:
Referring to the standard regulation of GB/T611, GB/T5009, GB/T13531, T22230, T24232, by adopting the buoyancy of the object immersed in the liquid, the density and volume of the irregular weight can be used as the testing standard of acid solution;
 
 
 
 
Specifications:
Model
TWD150 AS
Testing type
Specific Gravity (S.G), Baume( Bé°), Concentration (C%)
Weighing precision
0.001g
S.G range
0.0001~2.0000 (glass weight)
Bé°range
0.1~66.0 Bé°
C% range
0.1%~93.2%
Standard interface
RS-232
 
 
Functions:
1. It has the compensation function of air buoyancy and the correct function of weight volume thermal expansion coefficient and temperature deviation value.
2. The temperature deviation value of Baume degree can be set.
3. According to the specific gravity value of acid solution, and then by using non-linear method, concentration can be converted.
 
Features:
1. The body is cast integrally by using aluminum alloy, the structure is stable; the unstable screen problem caused by load cell measuring influenced by pressing the key and the damage problem caused by acid solution penetrating into circuit board and sensor will not occur.
2. The measuring position and sensor are on the same platform, it can improve the defect of the measuring deviation caused by corner load.
3. The screen independent design is adopted and the screen is separated from the measuring platform, the problem of the screen damaged by acid solution corrosion and adhesion will not occur.
4. With the special hook design, and just 50CC, the acid solution S.G can be showed directly, and then Bé°, C% can be converted.
 
Key measuring points:
The best testing environment for acid solution is under the international standard reference temperature condition of 15.6℃, 20℃.
 
 

1、Liquid thermostatic device TW-01A & beaker extractor TP-28:  

 
 
 
  
 
 
 TP-28
 
 
 
Standard demonstration: please refer to catalog TW-01A 
 
 
 
 

2、Liquid thermostatic water cycling device TW-01B:

 
 
 
 
Standard demonstration: please refer to catalog TW-01B